083005A/ 083010A/ 083005D/ 083005MD/ 083010MD/ 083025MD/ 083060MD

-          For lab and field applications

 

-          Cable lengths from 1.5 to 20 meters

 

 

Cat. No.

083005A

083010A

083005D

083005MD  

083010MD

083025MD

083060MD

Cable Length

 

 

1.5 m

 

3 m

 

1.5 m

 

1.5 m

 

3 m

 

10 m

 

20 m

 

Meter

Compatibility

 

830A, 835A, 862A

 

830A, 835A, 862A

 

810Aplus,

850Aplus

 

Star Series

 

Star Series

 

Star Series

 

Star Series

 

Application

 

 

Laboratory /

Field

 

Laboratory /

Field

 

Laboratory /

Field

 

Laboratory /

Field

 

Laboratory /

Field

 

Laboratory /

Field

 

Laboratory /

Field

 

Response

Time

90% of final

value in 10 sec;

95% of final

value in 15 sec;

99% of final

value in 60 sec

 

99% of final

value in 60 sec

 

99% of final

value in 60 sec

 

99% of final

value in 60 sec

 

99% of final

value in 60 sec

 

Minimum

Sample Flow

 

10 cm/sec

 

10 cm/sec

 

10 cm/sec

 

10 cm/sec

 

10 cm/sec

 

10 cm/sec

 

10 cm/sec

 

Oxygen

Consumption

 

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

0.008 µg/h (mg/L)-1

at 20 °C

Maximum

Allowed

Overpressure

6 BAR

 

6 BAR

 

6 BAR

 

6 BAR

 

6 BAR

 

6 BAR

 

6 BAR

 

Sample

Temperature

 

0 to 50 °C

 

0 to 50 °C

 

0 to 50 °C

 

0 to 50 °C

 

0 to 50 °C

 

0 to 50 °C

 

0 to 50 °C

 

Temperature

Sensor

Dual thermally

separated

Dual thermally

separated

Dual thermally

separated

Dual thermally

separated

Dual thermally

separated

Dual thermally

separated

Dual thermally

separated

Drift

 

 

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Approx. 0.1%

per day

 

Electrolyte

Lifetime

 

 

180 days

 

180 days

 

180 days

 

180 days

 

180 days

 

180 days

 

180 days

 

Storage

Temperature

 

-5 to 50 °C

 

-5 to 50 °C

 

-5 to 50 °C

 

-5 to 50 °C

 

-5 to 50 °C

 

-5 to 50 °C

 

-5 to 50 °C

 

Membrane

Caps

 

Membrane

Cap 080515

 

Membrane

Cap 080515

 

Solution

080514

 

Solution

080514

 

Solution

080514

 

Solution

080514

 

Solution

080514

 

Maintenance

Kit

 

080513

 

080513

 

080513

 

080513

 

080513

 

080513

 

080513

 

 

 

 

 

 

 

 