083150A

  • For field applications

 

  • Includes probe guard and 50 meter cable

 

Cat. No.

 

 

083150A

Cable

Length

 

50 m

Meter

Compatibility

 

830A, 835A

Application

 

 

Field

Response

Time

 

90% of final

value in 10 sec;

95% of final

value in 18 sec;

99% of final

value in 60 sec

Minimum

Sample Flow

 

20 cm/sec

Oxygen

Consumption

 

0.06 μg/h (mg/L)-1

at 20 ˚C

Maximum

Allowed

Overpressure

10 BAR

Sample

Temperature

 

0 to 50 ˚C

Temperature

Sensor

 

Dual,

thermally

separated

Drift

 

 

< 0.3% per

day

Electrolyte

Lifetime

 

180 days

Storage

Temperature

 

5 to 50 ˚C

Membrane

Caps

 

Membrane

Cap? 080515

Solution?

080514

Maintenance

Kit

 

080513

 

Ankersmid Lab          info@ankersmid-lab.nl,  (31)-0548-59 59 79